Pressure and flow in the extracorporeal circuit

Authors
Citation
Hd. Polaschegg, Pressure and flow in the extracorporeal circuit, CLIN NEPHR, 53, 2000, pp. S50-S55
Citations number
13
Categorie Soggetti
Urology & Nephrology","da verificare
Journal title
CLINICAL NEPHROLOGY
ISSN journal
03010430 → ACNP
Volume
53
Year of publication
2000
Supplement
1
Pages
S50 - S55
Database
ISI
SICI code
0301-0430(200002)53:<S50:PAFITE>2.0.ZU;2-8
Abstract
The impact of pressure and now changes in the extracorporeal circuit on dia lysis efficiency in response to altered hematocrit is discussed in the ligh t of the known properties of blood, published studies and the author's as y et unpublished measurements. Increasing the hematocrit from 0.35 to 0.45 mu st be expected to reduce the efficiency of urea elimination by 3 - 5%, whic h can be offset by increasing dialysis time or blood now. The change in phy sical parameters thus represents no obstacle to normalizing the hematocrit. However, basing the erythropoietin dose on the predialysis hematocrit shou ld be reconsidered.