Phase and light shift determination in an optically pumped cesium beam frequency standard

Citation
A. Makdissi et al., Phase and light shift determination in an optically pumped cesium beam frequency standard, IEEE ULTRAS, 47(2), 2000, pp. 461-465
Citations number
19
Categorie Soggetti
Optics & Acoustics
Journal title
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL
ISSN journal
08853010 → ACNP
Volume
47
Issue
2
Year of publication
2000
Pages
461 - 465
Database
ISI
SICI code
0885-3010(200003)47:2<461:PALSDI>2.0.ZU;2-K
Abstract
The determination of the phase shift and the light shift in our optically p umped Cs beam is reported here. We show that an error on the phase differen ce value determined by our fit method, without beam reversal, can be induce d if the light shift is not taken into account. If the experimental data ar e corrected for the light shift, the results of the fit method are in very good agreement with the beam reversal results. This allows us to reduce the related uncertainty in the accuracy budget of our standard. The resulting overall accuracy of the standard is now estimated to be 6.3 x 10(-15).