Headspace gas chromatographic measurements of limiting activity coefficients of eleven alkanes in organic solvents at 25 degrees C. 2. Accuracy and precision
Cb. Castells et al., Headspace gas chromatographic measurements of limiting activity coefficients of eleven alkanes in organic solvents at 25 degrees C. 2. Accuracy and precision, J CHEM EN D, 45(2), 2000, pp. 376-381
The accuracy of the infinite dilution activity coefficients reported in Par
t 1 is examined here through two independent tests. First, the data are cri
tically compared to values reported in the literature which were obtained b
y several different techniques, including headspace gas chromatography. Sec
ond, the data are validated by studying the consistency between our partiti
on coefficient data for normal alkanes (n-pentane to n-nonane) and those ob
tained for the solutes methane to butane by completely different methods. A
long-term study performed by using n-hexadecane as a solvent allowed us to
control the consistency of this set of measurements and also established t
hat the precision of the method is better than 5%.