T-1 Fast Acquisition Relaxation Mapping (T-1-FARM): Optimized data acquisition

Citation
Ca. Mckenzie et al., T-1 Fast Acquisition Relaxation Mapping (T-1-FARM): Optimized data acquisition, MAGN RES IM, 18(2), 2000, pp. 129-138
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
MAGNETIC RESONANCE IMAGING
ISSN journal
0730725X → ACNP
Volume
18
Issue
2
Year of publication
2000
Pages
129 - 138
Database
ISI
SICI code
0730-725X(200002)18:2<129:TFARM(>2.0.ZU;2-8
Abstract
A theoretical procedure for estimating the precision of the T-1 Fast Acquis ition Relaxation Mapping sequence as a function of a number of acquisition parameters has been validated by both simulations and experimental results. These results have clarified the selection of sequence parameters to give optimal accuracy and precision in the R-1* measurements. There is excellent agreement between theory, simulation, and experiment except for flip angle s greater than 9 degrees, at which point slice profile imperfections signif icantly degrade the precision of the technique. The experimental results in dicate that over a range of T(1)s that would be seen in a bolus tracking ex periment (25-1200 ms), T-1 Fast Acquisition Relaxation Mapping can be used to obtain 64 X 128 R-1* maps at a rate of 1 map/s, with a precision of 10% or better. (C) 2000 Elsevier Science Inc. All rights reserved.