Drift-corrected calibration in glow discharge optical emission spectrometry

Citation
Mr. Winchester et Jk. Miller, Drift-corrected calibration in glow discharge optical emission spectrometry, J ANAL ATOM, 15(2), 2000, pp. 169-175
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
ISSN journal
02679477 → ACNP
Volume
15
Issue
2
Year of publication
2000
Pages
169 - 175
Database
ISI
SICI code
0267-9477(200002)15:2<169:DCIGDO>2.0.ZU;2-N
Abstract
The effects of drifting instrument sensitivity in glow discharge optical em ission spectrometry (GD-OES) on instrument calibration were investigated. I t was found that for some analytical applications, the rate of drift may be fast relative to the time needed to calibrate the instrument. The results indicate that if this is the case, the validity of calibration may be compr omised, introducing analytical bias. Anew drift correction method recently published elsewhere was shown to eliminate any drift-induced calibration di fficulties and the associated potential for analytical error. This is impor tant,because drift correction methods that have been traditionally used in OES for decades are not applicable to this particular type of drift. Theref ore, the new drift correction method should enhance;the utility of GD-OES f or important applications, such as quality control and quality assurance in industry.