Fat suppression magnetic resonance imaging of the triangular fibrocartilage complex - Comparison with spin echo, gradient echo pulse sequences and histology
T. Nakamura et al., Fat suppression magnetic resonance imaging of the triangular fibrocartilage complex - Comparison with spin echo, gradient echo pulse sequences and histology, J HAND S-BR, 24B(1), 1999, pp. 22-26
Citations number
17
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine
Journal title
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME
Magnetic resonance imaging (MRI) of the triangular fibrocartilage complex (
TFCC) of the wrist was performed in ten healthy volunteers using spin echo
T1-weighted (SE-TI), fast spin echo T2-weighted (FSE-T2), gradient echo T2*
-weighted (GRE-T2*) and fat suppression spin echo T1-weighted (FS-T1) image
s. The images were obtained in the coronal plane and were compared to the c
orresponding histological coronal sections obtained from five fresh frozen
cadavers. In our analysis, the FS-T1 pulse sequence visualized the details
of the TFCC best, followed by the GRE-T2* images. Delineation of the TFCC o
n the SE-T1 and the FSE-T2 was poor. The coronal morphology of the TFCC rep
resented on the fat suppression image was almost identical to the correspon
ding histological sections. We conclude that the fat suppression MRI clearl
y shows the complex structure of the TFCC and is useful for the morphologic
al evaluation of the TFCC.