An overview of common photon beam quality specifiers used in radiotherapy d
osimetry introduces a reasoned discussion on the advantages and disadvantag
es of TPR20,10 and PDD(10)(x). It is shown that some of the potential advan
tages of PDD(10), are also present in other well known beam quality specifi
ers such as d(80) However, all PDD-based beam quality indices, including PD
D(10)(x), are subject to electron contamination and their determination is
affected by practical limitations. The proposed filtration of contaminant e
lectrons by Kosunen and Rogers [Med. Phys. 20, 1181-1188 (1993)] and by Li
and Rogers [Med. Phys. 21, 791-798 (1994)] is questioned, not only with reg
ard to the adequacy of using lead as an electron filter, but also in relati
on to its efficiency (if there were no contamination, restrictions for beam
calibrations at d(max) would be removed) and practical measurement. It is
argued that (i) there is no unique beam quality specifier that works satisf
actorily in all possible conditions, for the entire energy range of photon
energies used in radiotherapy and all possible accelerators used in hospita
ls and in standards laboratories, and (ii) TPR20,10 remains to be the most
appropriate specifier fur clinical photon beams as it has less practical dr
awbacks than PDD-based quality indices. The final impact on clinical photon
beam dosimetry resulting from the use of different photon beam quality spe
cifiers, is that they are not expected to yield a significant change (i.e.,
more than 0.5% and in most cases well within 0.2%) in the absorbed dose to
water in reference conditions for most clinical beams. (C) 2000 A,American
Association of Physicists in Medicine. [S0094-2405(00)00103-6].