Fm. Waterman et Ap. Dicker, Determination of the urethral dose in prostate brachytherapy when the urethra cannot be visualized in the postimplant CT scan, MED PHYS, 27(3), 2000, pp. 448-451
Citations number
8
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
This study investigates the feasibility of locating the urethra at the geom
etric center of peripherally loaded I-125 prostate implant when a urinary c
atheter is not utilized for the postimplant CT scan. Twenty postimplant CT
scans utilizing a urinary catheter were randomly selected. The urethra was
localized in each study and, in addition, a surrogate urethra was localized
at the geometric center of the prostate. Dose-volume histograms of the ure
thra and surrogate urethra were compiled and compared. The values obtained
for the urethra D-10, D-25, and D-50 were in good agreement and demonstrate
that the urethral dose can be determined reliably by locating a surrogate
urethra at the geometric center of the prostate in a peripherally loaded im
plant when the urethra cannot be visualized. (C) 2000 American Association
of Physicists in Medicine. [S0094-2405(00)01403-6].