Differential pulse polarographic determination of selenium(IV) in whole blood using the catalytic hydrogen wave

Citation
R. Inam et al., Differential pulse polarographic determination of selenium(IV) in whole blood using the catalytic hydrogen wave, TALANTA, 51(4), 2000, pp. 825-830
Citations number
25
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
51
Issue
4
Year of publication
2000
Pages
825 - 830
Database
ISI
SICI code
0039-9140(20000403)51:4<825:DPPDOS>2.0.ZU;2-6
Abstract
The selenium content in blood was determined using the hydrogen catalytic p eak. This peak at -1.1 V was obtained in the presence of selenium and molyb denum at pH values of 1-4 in different buffers. For the determination of se lenium, the Mo(VI) concentration has to be similar to 100-200 times higher than the selenium present. The linear domain range of selenium is 1 x 10(-6 )-5 x 10(-9) M. The interference of zinc is eliminated by the addition of E DTA at pH 3.5 acetate buffer. The method was applied to 1.0 mi of digested blood, and 620 +/- 44 mu g l(-1) Se and 7.15 mg l(-1) Zn could be determine d with a 90% (n = 6) confidence interval. (C) 1999 Elsevier Science B.V. Al l rights reserved.