Nm. Alves et Jf. Mano, Temperature correction of dynamic mechanical and thermomechanical analysers during heating, cooling and isothermal experiments, THERMOC ACT, 346(1-2), 2000, pp. 133-145
In this work some procedures are outlined to improve the correction of the
temperature axis in isochronal and isothermal experiments performed in DMA
and TMA apparatus. For the case of experiments on heating, the influence of
the position of the metallic standard used during the calibration procedur
es is found to have great importance, mainly due to the purge gas flux insi
de the furnace. New calibration routines are proposed, taking into account
the mechanical mode used. The isothermal correction can be performed by ext
rapolating heating results for zero scanning rate or by running experiments
at a very low heating rate. A method for correcting the temperature of exp
eriments on cooling with data obtained from heating experiments is also pro
posed, provided that symmetry between heat fluxes of cooling and heating ex
periments exists. This method was confirmed by investigating the temperatur
e position of the alpha-relaxation of a thermoset polymer, detected by DMA
experiments on heating and cooling. (C) 2000 Elsevier Science B.V. All righ
ts reserved.