FORCE MEASUREMENT ON FRACTURE SITE WITH EXTERNAL FIXATION

Citation
Zg. Wang et al., FORCE MEASUREMENT ON FRACTURE SITE WITH EXTERNAL FIXATION, Medical & biological engineering & computing, 35(3), 1997, pp. 289-290
Citations number
3
Categorie Soggetti
Engineering, Biomedical","Computer Science Interdisciplinary Applications","Medical Informatics
ISSN journal
01400118
Volume
35
Issue
3
Year of publication
1997
Pages
289 - 290
Database
ISI
SICI code
0140-0118(1997)35:3<289:FMOFSW>2.0.ZU;2-W
Abstract
A force measurement device has been designed to monitor the mechanical properties of fracture site with external fixation. Forces are measur ed through electric resistance strain gauges mounted on fixation frame work and the measurement results are displayed on an LCD screen. The d evice features a force range of 0-10 kg with linearity and repeatabili ty less than 1% and accuracy less than 0.1 kg.