The possibilities of computer-assisted and automated accelerator mass spect
rometry (AMS) measurements were explored. The goal of these efforts is to d
evelop fully automated procedures for "routine" measurements at the Vienna
Environmental Research Accelerator (VERA), a dedicated 3-MV Pelletron tande
m AMS facility. As a new tool for automatic tuning of the ion optics we dev
eloped a multi-dimensional optimization algorithm robust to noise, which wa
s applied for C-14 and Be-10. The actual isotope ratio measurements are per
formed in a fully automated fashion and do not require the presence of an o
perator. Incoming data are evaluated online and the results can be accessed
via Internet. The system was used for C-14,Be-10,Al-26 and I-129 measureme
nts. (C) 2000 Elsevier Science B.V. All rights reserved.