Calibration of temperature modulated calorimeters (TMC) is essential becaus
e it is often not possible to calculate heat capacity and sample temperatur
e directly from the quantities measured. The question arises whether temper
ature calibration used in conventional calorimetry is also valid for TMC's.
To prove this, a well defined transition is needed which does not disturb
the temperature profile within the sample during modulation. The response o
f the whole system (calorimeter, sample, transition) must be linear. Some l
ow energy liquid crystal transitions behave in this way, under certain cond
itions, and can be used for temperature calibration in TMC's. The procedure
is demonstrated for the nematic to smectic-A transition in a cyanobiphenyl
liquid crystal (80CB). The applicability of 80CB for temperature calibrati
on is shown for temperature modulated DSC, AC calorimeter, 3 omega method a
nd photo-acoustic method. (C) 2000 Elsevier Science B.V. All rights reserve
d.