Abstract
Coulometric nanotitrations were realized in a microchannel system using a c ontinuous-flow titration technique with a triangle current-time profile. Re dox and acid-base titrations were carried out on Fe(II) and nitric acid sam ples, respectively, with the same nanotitrator device. A linear relation be tween the concentration and the coulometric current transferred to the solu tion was found. The advantages of this universally applicable nanotitrator are fast response, low sample volume, high sensitivity, and high reproducib ility as well as the convenience of handling an automated analyzer of the n ow-through type.