Qg. Wang et al., Capillary electrophoresis based immunoassay for monoclonal antibody with diode laser induced fluorescence detection, ANAL LETTER, 33(4), 2000, pp. 589-602
A capillary electrophoresis based immunoassay (CEIA) for monoclonal antibod
y using diode laser induced fluorescence (LIF) detection was described. A d
irect assay for monoclonal anti-BSA in mouse serum was used as a model. BSA
was labeled with Cy5 and used as the immunoreagent. The 635 nm line of a d
iode laser was used as the excitation source for LIF detection. The calibra
tion curve for anti-BSA in mouse serum had a linear dynamic range of 4-40 n
M. The concentration limit of detection (LOD) was 1.2 nM. Incubation time a
nd CE conditions such as buffer concentration, pH and separation voltage we
re optimized, and the performances of different lasers as excitation source
s were also compared.