Incidents involving implantable catheter ports, in 1996/1998: analysis of the data base of the French Health Ministry.

Citation
F. Conreux et al., Incidents involving implantable catheter ports, in 1996/1998: analysis of the data base of the French Health Ministry., ANN FR A R, 19(3), 2000, pp. 171-176
Citations number
19
Categorie Soggetti
Aneshtesia & Intensive Care
Journal title
ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION
ISSN journal
07507658 → ACNP
Volume
19
Issue
3
Year of publication
2000
Pages
171 - 176
Database
ISI
SICI code
0750-7658(200003)19:3<171:IIICPI>2.0.ZU;2-V
Abstract
We analyse all incident reports to the French Health Ministry over the 1996 -1998 period, concerning implantable catheter ports. They represent 7% of a ll reports in the field of anaesthesia and intensive care. Two hundred elev en reports have been analysed. There were 93 catheter ruptures with 77 intr avascular migrations. In more than 50% of cases, a pinch-off syndrome mecha nism was responsible of these ruptures and a displacement of the O'ring was involved in 25%. Additional minor complications were reported and discusse d in regard to literature. We conclude that implantable catheter ports may lead to serious incidents which are preventable in most cases. (C) 2000 Edi tions scientifiques et medicales Elsevier SAS.