An optical-electronic method was developed for measurement of falling liqui
d film thicknesses. The method is based on the postulate that a sheet light
beam passing tangentially through a vertical liquid film on a cylinder wil
l be blocked by the falling liquid film. Hence, when the beam is much wider
than the film, the output of a photodiode probe which is located on the op
posite side of the cylinder from the light source will be reduced by an amo
unt proportional to the film thickness. The shadow of the transient film sh
ape will also appear on a screen behind the falling film. Therefore, the va
riation of the amplitude of the film waves can be measured from the output
of the photodiode and the average film thickness can be measured from pictu
res of the shadow using computer aided image analysis. The average film thi
ckness measured agrees well with commonly accepted empirical equations. The
maximum relative deviation between the experimental and calculated results
is 18.5% with a standard deviation of 4.34 x 10(-2) mm. Compared with popu
lar conductance parallel-wire probes, this method has the advantages of con
venience, noninvasive, higher spatial and temporal resolution, smaller hyst
eresis and it does not disturb the model's surface. (C) 2000 American Insti
tute of Physics. [S0034-6748(00)00504-9].