Resolution of a position-sensitive proportional-counter with a resistive anode wire of carbon fiber

Citation
D. Ohsawa et al., Resolution of a position-sensitive proportional-counter with a resistive anode wire of carbon fiber, APPL RAD IS, 52(4), 2000, pp. 943-954
Citations number
47
Categorie Soggetti
Multidisciplinary
Journal title
APPLIED RADIATION AND ISOTOPES
ISSN journal
09698043 → ACNP
Volume
52
Issue
4
Year of publication
2000
Pages
943 - 954
Database
ISI
SICI code
0969-8043(200004)52:4<943:ROAPPW>2.0.ZU;2-9
Abstract
Theoretical and experimental analyses have been performed to examine the po sition resolution of a proportional counter with a 7-mu m diameter carbon f iber as a resistive anode, which is used to encode one-dimensional position s of incident photons according to the charge-division method. These analys es have clearly shown that the carbon fiber is applicable to the position s ensing of the order of 100-200 mu m as long as the filled gas is pressurize d at near 10 atm. (C) 2000 Elsevier Science Ltd. All rights reserved.