Reaming increases intramedullary pressure and can cause fat intravasation.
This study investigated the influence of reaming parameters on pressure cha
nges using the AO, Gray, and Howmedica reamers. Silicone cylindrical shells
with 10-, 11-, 16-, 19-, and 20-mm inner diameters were filled with a 1:2
ratio of paraffin and petroleum jelly. The 10- and 11-mm shells were reamed
using a 9-mm reamer, the 16-mm shells with a 14-mm reamer, and the 19- and
20-mm shells with an 18-mm reamer. Each experiment was performed seven tim
es while singularly varying driving speeds of 15, 30, and 50 mm per second,
with a revolution rate of 150, 250, and 500 revolutions per minute. Using
stepwise linear regression analysis, driving speed was the most significant
parameter. The lowest pressure increases occurred at the lowest driving sp
eed and the highest revolution rate in all cylinders. The model used for th
is study did not take into consideration the interaction between the cortic
al shell and reamer head. The results of this study suggest that reaming sh
ould be done at a low driving speed and a high revolution rate using a smal
l cored reamer head, especially during the first reaming steps.