The behaviour of fine-mesh vacuum phototriodes (VPTs) with the external dia
meters of 21 and 25 mm has been investigated in an axial magnetic field up
to 4 T in view of their applications as readout devices for CMS Endcap Elec
tromagnetic Calorimeter. The measured VPT parameters are: the photocathode'
s sensitivity and its homogeneity, the gain in zero and 4 T magnetic field
at tilt angles corresponding to the pseudorapidity range of CMS ECAL Endcap
1.48-3.0 as a function of fine-mesh cell dimensions, excess noise factor a
nd the stability of the photocathode response under the illumination by lig
ht emission diodes (LED) and the irradiation by 14 MeV neutrons. Phototriod
es with 100 lines per mm fine-mesh and 25 mm external diameter are found to
be the best candidates for coupling with rear PbWO4 crystals by dimensions
of 30 x 30 mm, proposed to be used in CMS ECAL Endcaps. VPTs provide a gai
n of the order (6-8) in a 4 T magnetic field and an excess noise factor of
2-2.5 under illumination of a full photocathode's area. (C) 2000 Elsevier S
cience B.V. All rights reserved.