Measurement of the dynamic response of low-gain solid-state photodetector under weak pulse illumination

Citation
D. Clement et al., Measurement of the dynamic response of low-gain solid-state photodetector under weak pulse illumination, NUCL INST A, 442(1-3), 2000, pp. 378-383
Citations number
5
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
442
Issue
1-3
Year of publication
2000
Pages
378 - 383
Database
ISI
SICI code
0168-9002(20000311)442:1-3<378:MOTDRO>2.0.ZU;2-X
Abstract
This paper presents the development of a Tunable Light Pulse Generator buil t to measure the quantum efficiency of photodetectors under continuous or p ulsed (20 ns FWHM) illumination as a function of wavelength in the Visible- UV region. For this purpose, a miniature Hamamatsu R5600U-04 PMT with low c athode resistivity can be used as reference detector. Dedicated electronics was developed in order to measure simultaneously the cathode and anode cur rents. This should allow to determine precisely the gain as a function of t he high voltage applied to the PMT. Preliminary measurements using the TLPG do not demonstrate significant differences between the dynamic (similar to 10 000 photons) and DC responses of PIN photodiodes down to 300 nm. (C) 20 00 Elsevier Science B.V. All rights reserved.