Background. A new children's hospital provided the impetus to investigate r
adiation dose and image quality in a fluoroscope that was specially enginee
red for pediatric fluoroscopy. Radiation protection management recommends r
adiation exposures that are as low as reasonably achievable, while still ma
intaining diagnostic image quality.
Objectives. To obtain comparative phantom imaging data on radiation exposur
e and image quality from a newly installed fluoroscope before and after opt
imization for pediatric imaging.
Materials and methods. Images were acquired from various thickness phantoms
, simulating differing patient sizes. The images were evaluated for visuali
zation of high- and low-contrast objects and for radiation exposure. Effect
s due to use of the image intensifier anti-scatter grid were also investiga
ted.
Results. The optimization of the new fluoroscope for pediatric operation re
duced radiation exposure by about 50% (compared to the originally installed
fluoroscope), with very little loss of image quality. Pulsed fluoroscopy w
as able to lower radiation dose to less than 10% of continuous fluoroscopy,
while still maintaining acceptable phantom image quality.
Conclusion. Radiation exposure in pediatric fluoroscopy can be reduced to v
alues well below the exposure settings that are typically found on unoptimi
zed fluoroscopes. Pulsed fluoroscopy is considered a requisite for optimal
pediatric fluoroscopy.