Coulometric micro-titrator with a ruthenium dioxide pH-electrode

Citation
C. Colombo et al., Coulometric micro-titrator with a ruthenium dioxide pH-electrode, ANALYT CHIM, 412(1-2), 2000, pp. 69-75
Citations number
20
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
412
Issue
1-2
Year of publication
2000
Pages
69 - 75
Database
ISI
SICI code
0003-2670(20000508)412:1-2<69:CMWARD>2.0.ZU;2-E
Abstract
A coulometric diffusional titrator with a ruthenium dioxide pH-sensing elec trode for the end-point detection in acid-base titrations is presented. The device was prepared by galvanic deposition of noble metals onto electrodes etched on a copper clad printed circuit board. Five electrodes were employ ed: three gold electrodes (two actuator and one counter) for the coulometri c production of the titrating protons or hydroxide ions, and a RuO2-electro de and Ag/AgCl pseudo-reference electrode pair for the end-point detection. The RuO2-electrode for pH-measurements showed a linear response to the var iation of the pH-value in the range from 2 to 12 pH-units and was useable f or a period of 3 months. The device was found to yield linear relationships between the analyte concentration and the equivalence time for concentrati on ranges of acetic acid and ammonia between about 0.1 and 100 mM. (C) 2000 Elsevier Science B.V. All rights reserved.