G. Totschnig et al., Multiplexed continuous-wave diode-laser cavity ringdown measurements of multiple species, APPL OPTICS, 39(12), 2000, pp. 2009-2016
Rapid cavity ringdown measurements of multiple broadband absorbing species
(methanol and isopropanol) in gas mixtures have been recorded with two mult
iplexed continuous-wave distributed-feedback diode lasers operating near 1.
4 mu m. A measurement sensitivity of 2.4 x 10(-9) cm(-1) for a 4.3-s averag
ing time was achieved in a 39.5-cm-long static cell with 99.94% reflectivit
y mirrors. This corresponds to a water-vapor detection limit of less than 2
ppb (parts in 10(9)) for the strong H2O lines near 1.4 mu m. The shot-to-s
hot noise of the decay time constant tau was approximately 0.3-0.7%, and ri
ngdown acquisition rates as great as 900 Hz were achieved. (C) 2000 Optical
Society of America.