S. Hartter et al., Automated determination of reboxetine by high-performance liquid chromatography with column-switching and ultraviolet detection, J CHROMAT B, 740(1), 2000, pp. 135-140
A fully automated method including column-switching and isocratic high-perf
ormance Liquid chromatography (HPLC) was developed for quantitative analysi
s of the new antidepressant reboxetine, a noradrenaline reuptake inhibitor.
After serum injection into the HPLC system and on-line sample clean-up on
a silica C-8 (10x4.0 mm I.D.) clean-up column with an eluent consisting of
2.5% acetonitrile in deionized water, the chromatographic separation was pe
rformed on an analytical column (Lichrospher CN; 250x4.6 mm I.D.) with an e
luent of acetonitrile-aqueous potassium phosphate buffer (0.008 M, pH 6.4)
(50:50). The UV detector was set at 273 or 226 nm. The limit of quantificat
ion was about 15 ng/ml at 273 nm and about 4 ng/ml at 226 nm. The day-to-da
y relative standard deviation ranged between 2.7 and 6.7% with recovery rat
es greater than or equal to 90%. Linear regression analyses revealed correl
ation coefficients >0.998. The method can be applied to therapeutic drug mo
nitoring of reboxetine as well as pharmacokinetic studies. (C) 2000 Elsevie
r Science B. V. All rights reserved.