S. Cavalcanti et al., Validation of automated oscillometric sphygmomanometer (HDBPM) for arterial pressure measurement during haemodialysis, MED BIO E C, 38(1), 2000, pp. 98-101
An HDBPM oscillometric sphygmomanometer used for the automatic measurement
of arterial blood pressure is evaluated according to the ANSI/AAMI SP10-199
2 standard. The accuracy of the HDBPM is ascertained by comparing it agains
t the standard Riva-Rocci ascultatory method. Following the ascultatory met
hod, two independent observers use the HDBPM devise to simultaneously measu
re the arterial blood pressure in 92 subjects of varying ages and having di
fferent blood pressures and arm sires. High agreement is found when compari
ng the observers' pressure determinations (within 10 mmHg for 100% of obser
vations). Correlation between the average of two ascultatory determinations
and the HDBPM is high both for the systolic (r = 0.98) and diastolic (r =
0.94) pressures. The mean of the differences between the pressures measured
by the observers and the HDBPM device are 0.2mmHg (systolic) and -0.4mmHg
(diastolic). The percentages of readings within 10mmHg between those taken
by the observers and those taken by the HDBPM are 88% (systolic) and 97% (d
iastolic). These results largely satisfy current requirements.