M. Bambynek et al., A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy, MED PHYS, 27(4), 2000, pp. 662-667
Citations number
28
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
A fast dosimetry system based on plastic scintillator detectors has been de
veloped which allows three-dimensional measurement of the radiation field i
n water of beta-sources appropriate for application in cardiovascular brach
ytherapy. This system fulfills the AAPM Task Group 60 recommendations for d
osimetry of cardiovascular brachytherapy sources. To demonstrate the use of
the system, measurements have been performed with an Y-90-wire source. The
dose distribution was determined with a spatial resolution of better than
0.2 mm, with only a few minutes needed per scan. The scintillator dosemeter
was absolutely calibrated in terms of absorbed dose to water with a precis
ion of +/-7.5%. The relative precision achievable is +/-2.5%. The response
of the system is linear within +/-2% for dose rates from 0.5 mGy s(-1) to 5
00 mGy s(-1). (C) 2000 American Association of Physicists in Medicine.