A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy

Citation
M. Bambynek et al., A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy, MED PHYS, 27(4), 2000, pp. 662-667
Citations number
28
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
27
Issue
4
Year of publication
2000
Pages
662 - 667
Database
ISI
SICI code
0094-2405(200004)27:4<662:AHHAFD>2.0.ZU;2-W
Abstract
A fast dosimetry system based on plastic scintillator detectors has been de veloped which allows three-dimensional measurement of the radiation field i n water of beta-sources appropriate for application in cardiovascular brach ytherapy. This system fulfills the AAPM Task Group 60 recommendations for d osimetry of cardiovascular brachytherapy sources. To demonstrate the use of the system, measurements have been performed with an Y-90-wire source. The dose distribution was determined with a spatial resolution of better than 0.2 mm, with only a few minutes needed per scan. The scintillator dosemeter was absolutely calibrated in terms of absorbed dose to water with a precis ion of +/-7.5%. The relative precision achievable is +/-2.5%. The response of the system is linear within +/-2% for dose rates from 0.5 mGy s(-1) to 5 00 mGy s(-1). (C) 2000 American Association of Physicists in Medicine.