CT number variations due to different image acquisition and reconstructionparameters: a thorax phantom study

Citation
R. Groell et al., CT number variations due to different image acquisition and reconstructionparameters: a thorax phantom study, COMP MED IM, 24(2), 2000, pp. 53-58
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS
ISSN journal
08956111 → ACNP
Volume
24
Issue
2
Year of publication
2000
Pages
53 - 58
Database
ISI
SICI code
0895-6111(200003/04)24:2<53:CNVDTD>2.0.ZU;2-H
Abstract
mA humanoid thorax phantom containing six compartments was scanned with two different computed tomography (CT) scanners using various image acquisitio n and reconstruction parameters. The differences of CT numbers were statist ically significant between the two CT scanners for each compartment (p < 0. 001) except for the "air" compartment. The variabilities of the CT numbers are described for the different parameters. The mean CT numbers of the "wat er" compartment, for instance, ranged from 1 to 15 HU (Hounsfield Units), t hose of the "air" compartment varied from -962 to -990 HU. Knowledge of the se CT number variabilities is necessary when CT numbers are used for tissue characterization. (C) 2000 Elsevier Science Ltd. All rights reserved.