R. Groell et al., CT number variations due to different image acquisition and reconstructionparameters: a thorax phantom study, COMP MED IM, 24(2), 2000, pp. 53-58
mA humanoid thorax phantom containing six compartments was scanned with two
different computed tomography (CT) scanners using various image acquisitio
n and reconstruction parameters. The differences of CT numbers were statist
ically significant between the two CT scanners for each compartment (p < 0.
001) except for the "air" compartment. The variabilities of the CT numbers
are described for the different parameters. The mean CT numbers of the "wat
er" compartment, for instance, ranged from 1 to 15 HU (Hounsfield Units), t
hose of the "air" compartment varied from -962 to -990 HU. Knowledge of the
se CT number variabilities is necessary when CT numbers are used for tissue
characterization. (C) 2000 Elsevier Science Ltd. All rights reserved.