Xj. Yang et al., Determination of Guan-Fu Base A, a new anti-arrhythmic, in human plasma bygas chromatography and electron-capture detection, J CHROMAT B, 740(2), 2000, pp. 273-279
A gas chromatographic-electron capture detection (GC-ECD) method has been d
eveloped for determining Guan-Fu Base A (GFA), an experimental anti-arrhyth
mic, in human plasma. The method was based on one-step liquid-liquid extrac
tion with toluene and chemical derivatization with pentafluoropropionic anh
ydride followed by GC-ECD. The derivatives of GFA and metoprolol (Met, inte
rnal standard) were confirmed by gas chromatography-mass spectrometry (GC-M
S) to be dipentafluoropropionyl-GFA and dipentafluoropropionyl-Met. The met
hod was linear over the concentration ranges of 0.1-20.0 and 1.0-30.0 mu g/
ml with the detection limit of 0.05 mu g/ml at S/N=5. The intra- and inter-
assay precisions were less than 6 and 10%, and accuracy 99.70+/-3.30 and 97
.60+/-5.99%, respectively. The absolute recoveries were 81.88, 77.35, 80.79
and 83.85% for GFA at concentrations of 0.5, 1.0, 5.0 and 14.0 mu g/ml and
88.24% for Met at 3.0 mu g/ml, respectively. (C) 2000 Elsevier Science B.V
. All rights reserved.