CF-LIPS: A new approach to LIPS spectra analysis

Citation
A. Ciucci et al., CF-LIPS: A new approach to LIPS spectra analysis, LASER PART, 17(4), 1999, pp. 793-797
Citations number
14
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
LASER AND PARTICLE BEAMS
ISSN journal
02630346 → ACNP
Volume
17
Issue
4
Year of publication
1999
Pages
793 - 797
Database
ISI
SICI code
0263-0346(1999)17:4<793:CANATL>2.0.ZU;2-H
Abstract
The usual technique of the laser induced plasma spectroscopy (LIPS) spectra analysis relies on the use of calibration curves for quantitative measurem ents of plasma composition. However, LIPS calibration curves strongly depen d on the material composition (the so-called matrix effect); thus, this sta ndard approach to the LIPS spectra analysis is inadequate when precise info rmation on unknown samples composition is required. In this paper we presen t a new procedure, based on LIPS, for calibration-free quantitative element al analysis of materials. The new procedure, based on an algorithm develope d and patented by IFAM-CNR, allows the overcoming of the matrix effect, yie lding precise and accurate quantitative results without the need of calibra tion curves.