Bj. Gerbi et Fm. Khan, PLANE-PARALLEL IONIZATION-CHAMBER RESPONSE IN THE BUILDUP REGION OF OBLIQUELY INCIDENT PHOTON BEAMS, Medical physics, 24(6), 1997, pp. 873-878
Fixed-separation plane-parallel ionization chambers have been shown to
overestimate the dose in the buildup region of normally incident high
-energy photon beams. This work shows that these ionization chambers e
xhibit an even greater over-response in the buildup region of obliquel
y incident photon beams. This over-response at oblique incidence is gr
eatest at the surface of the phantom and increases with increasing ang
le of beam incidence. In addition, the magnitude of the over-response
depends on held size, beam energy, and chamber construction. This stud
y shows that plane-parallel ionization chambers can over-respond by mo
re than a factor of 2.3 at the phantom surface for obliquely incident
high-energy photon fields. (C) 1997 American Association of Physicists
in Medicine.