Performance of multiplexed SQUID readout for Cryogenic Sensor Arrays

Citation
Ja. Chervenak et al., Performance of multiplexed SQUID readout for Cryogenic Sensor Arrays, NUCL INST A, 444(1-2), 2000, pp. 107-110
Citations number
3
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
444
Issue
1-2
Year of publication
2000
Pages
107 - 110
Database
ISI
SICI code
0168-9002(20000411)444:1-2<107:POMSRF>2.0.ZU;2-4
Abstract
We report on the implementation of a multiplexer that uses superconducting quantum interference devices (SQUIDs) to read out low-impedance cryogenic d etectors. Using prototype chips, a circuit was built which interfaces eight input SQUID channels with a close-packed array of eight transition-edge se nsor (TES) infrared bolometers. Circuit elements were measured and crosstal k specifications are reported. Digital feedback is employed to flux-lock a single element in the array of SQUIDs. Published by Elsevier Science B.V.