The microcalorimeter spectrometer on the ASTRO-E X-ray observatory

Citation
Rl. Kelley et al., The microcalorimeter spectrometer on the ASTRO-E X-ray observatory, NUCL INST A, 444(1-2), 2000, pp. 170-174
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
444
Issue
1-2
Year of publication
2000
Pages
170 - 174
Database
ISI
SICI code
0168-9002(20000411)444:1-2<170:TMSOTA>2.0.ZU;2-B
Abstract
The Astro-E Observatory will employ a high-resolution X-ray spectrometer ba sed on a 32 pixel array of microcalorimeters with an energy resolution of a bout 12 eV at 6 keV (the Fe-K region). This will provide spectral resolving power 10 times higher than any other non-dispersive X-ray spectrometer. Th e instrument incorporates a three-stage cooling system capable of maintaini ng the temperature of the detector stage at 60 mK for about two years in or bit with a 97% duty cycle. The array sits at the focus of a grazing inciden ce conical mirror. The quantum efficiency of the microcalorimeters and the reflectivity of the X-ray mirror system combine to give high throughput ove r the 0.3-12 keV energy band. This new capability will enable the study of a wide range of high-energy astrophysical sources with unprecedented spectr al sensitivity. (C) 2000 Published by Elsevier Science B.V. All rights rese rved.