Design of QET phonon sensors for the CDMS ZIP detectors

Citation
T. Saab et al., Design of QET phonon sensors for the CDMS ZIP detectors, NUCL INST A, 444(1-2), 2000, pp. 300-303
Citations number
5
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
444
Issue
1-2
Year of publication
2000
Pages
300 - 303
Database
ISI
SICI code
0168-9002(20000411)444:1-2<300:DOQPSF>2.0.ZU;2-P
Abstract
The Cryogenic Dark Matter Search (CDMS) ZIP detectors utilize quasiparticle trapping as the mechanism for coupling the energy of a particle interactio n in the Ge (or Si) absorber into a tungsten (W) transition edge sensor (TE S). Consequently, the dynamics of quasiparticle propagation and loss signif icantly impact the energy collection and resolution of the detector. This p aper describes the considerations necessary in optimizing the detector surf ace geometry in order to have maximal quasiparticle collection. (C) 2000 El sevier Science B.V. All rights reserved.