Simultaneous measurement of phonons and scintillation light for active background rejection in the CRESST experiment

Citation
M. Bravin et al., Simultaneous measurement of phonons and scintillation light for active background rejection in the CRESST experiment, NUCL INST A, 444(1-2), 2000, pp. 323-326
Citations number
7
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
444
Issue
1-2
Year of publication
2000
Pages
323 - 326
Database
ISI
SICI code
0168-9002(20000411)444:1-2<323:SMOPAS>2.0.ZU;2-P
Abstract
We report on simultaneous detection of phonons and scintillation light usin g a cryogenic calorimeter composed of a scintillating CaWO4 crystal and a s eparate light detector, both instrumented with a tungsten superconducting p hase transition thermometer operating at about 12 mK. Clear discrimination between electron and nuclear recoils was demonstrated by irradiating the de tector with electrons, photons and neutrons. The shape of the nuclear recoi l energy spectrum induced by the neutrons could be well reproduced by simul ations. (C) 2000 Elsevier Science B.V. All rights reserved.