Xd. Lin et Jk. Xie, Space-time resolving vacuum ultraviolet spectrometer based on a rotating polyhedral mirror, REV SCI INS, 71(5), 2000, pp. 2068-2070
Using a rotating polyhedral mirror and a vacuum ultraviolet (VUV) monochrom
ater, a space-time resolving VUV diagnostic system is developed. Measuremen
t of the O VI (103.2 nm) radiation on the HT-6M tokamak shows that the time
resolution of the system is better than 4 ms and the space resolution is b
etter than 2 cm. Compared with traditional instruments, this system has imp
roved measurement efficiency, and error from shot-to-shot discharge variati
ons is avoided. (C) 2000 American Institute of Physics. [S0034-6748(00)0110
5-9].