S. Ciolac et E. Vasilescu, The corrosion behaviour of some titanium alloys in physiological environment. Short term "in vitro" studies, REV RO CHIM, 44(5), 1999, pp. 431-438
Short term tests taken on Ti-5Al-2.5Fe and Ti-6Al-4V alloys in physiologica
l environments (0.9% NaCl and Ringer's solution with adjusted pH) have reve
aled that there is no risk of localised corrosion (pitting or crevice) as a
consequence of non-uniform composition of the electrolyte along the metall
ic surface.