Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition

Citation
H. Shin et al., Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition, EUR J CAR-T, 17(4), 2000, pp. 492-494
Citations number
6
Categorie Soggetti
Cardiovascular & Respiratory Systems
Journal title
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
ISSN journal
10107940 → ACNP
Volume
17
Issue
4
Year of publication
2000
Pages
492 - 494
Database
ISI
SICI code
1010-7940(200004)17:4<492:AIHFRF>2.0.ZU;2-L
Abstract
We describe a rare intraaortic balloon pump (IABP) vascular complication as a result of malpositioning of the IABP. A 61-year-old man with unstable an gina underwent emergency coronary artery bypass grafting soon after the ins ertion of an IABP. postoperative hemodynamics were stable, but acute hepati c dysfunction occurred on the second postoperative day. Doppler echography revealed the absence of hepatic arterial Row. The IABP was removed, and art erial flow was immediately restored. Thereafter, the hepatic function recov ered rapidly. This is a rare case that demonstrates how IABP can cause mech anical abdominal arterial branch obstruction. Evaluations using Doppler ech ography are useful in detecting such IABP complications. (C) 2000 Elsevier Science B.V. All rights reserved.