The trace element analysis of superconductor oxides YBa2Cu3O7-x was perform
ed by use of a PU 7000 sequential inductively coupled plasma atomic emissio
n spectrometer (ICP-AES). The manganese emission line at 259.373 nm was use
d to find the optimum viewing position in the plasma for the determination
of Ca, Mg, Fe, Mn, Al, Ni, Si and Sr. The different emission lines were com
pared and matrix effects of major elements (Cu, Ba and Y) were studied. A s
can of emission lines Mn II at 259.373 nm, Mg II at 280.270 nm, Fe II at 25
9.940 nm, Al I at 309.271, Ni II at 221.647 nm, Ca II at 396.847 nm, Si I a
t 251.611 nm and Sr II at 407.771 nm was performed. The detection limits (D
L) of the elements were in the range 0.002-0.006 mg l(-1) and the practical
detection limits (PDL) were between 2.0 x 10(-5) and 1.2 x 10(-4) (% m/m).
The precision of the determination (RSD) ranged from 1.4 to 3.2% and was d
ependent on the element and its concentration.