Ll. Fang et al., Evaluation of evaporative light-scattering detector for combinatorial library quantitation by reversed phase HPLC, J COMB CHEM, 2(3), 2000, pp. 254-257
A quantitation study using reversed phase HPLC with UV and evaporative ligh
t-scattering detector (ELSD) was conducted on 90 library standards selected
from 15 small molecule combinatorial libraries (six standards from each li
brary). This study assessed the quantitation errors using a single calibrat
ion curve for rapid purity analysis of combinatorial libraries. The average
quantitation error of six standards from one library at 200 mu M by UV was
13.4%, 20.6%, and 60.3%, at 214, 220, and 254 nm, respectively. By ELSD, t
he average quantitation error of these six standards at 200 mu M was only 7
.7%. Applying this ELSD calibration curve to 84 standards from 14 structura
lly diverse libraries, an average quantitation error of 16.4% was obtained.
The average quantitation error of all 90 standards from 15 libraries using
15 calibration curves was 18.5%.