Calculated dose response of Gafchromic MD55 film for Pd-103 and I-125 relative to Co-60

Citation
Wv. Prestwich et Rj. Murphy, Calculated dose response of Gafchromic MD55 film for Pd-103 and I-125 relative to Co-60, RADIAT MEAS, 32(3), 2000, pp. 173-179
Citations number
12
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
RADIATION MEASUREMENTS
ISSN journal
13504487 → ACNP
Volume
32
Issue
3
Year of publication
2000
Pages
173 - 179
Database
ISI
SICI code
1350-4487(200006)32:3<173:CDROGM>2.0.ZU;2-8
Abstract
Generalized cavity theory is used to calculate the dose-response of the rad iochromic dye film MD55 to the X- and gamma-radiations from Pd-103 and I-12 5. Th, analysis indicates that for film calibrated using Co-60, the correct ion factor giving the dose to water is 1.81 +/- 0.05 for Pd-103 and 1.71 +/ - 0.05 for I-125. (C) 2000 Elsevier Science Ltd. All rights reserved.