Development of a potentiometric flow cell with a stainless steel electrodefor pH measurements. Determination of acid mixtures using flow injection analysis

Citation
Cg. Zampronio et al., Development of a potentiometric flow cell with a stainless steel electrodefor pH measurements. Determination of acid mixtures using flow injection analysis, TALANTA, 51(6), 2000, pp. 1163-1169
Citations number
13
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
51
Issue
6
Year of publication
2000
Pages
1163 - 1169
Database
ISI
SICI code
0039-9140(20000505)51:6<1163:DOAPFC>2.0.ZU;2-0
Abstract
In this work a stainless steel electrode was prepared, characterised and us ed as indicator electrode in a potentiometric flow cell. Due to its versati lity, it was possible to study different electrode geometries and flow cell arrangements. After optimising the system, mixtures of succinic and oxalic acids were determined by titration. Partial least squares (PLS) regression as multivariate calibration tool was applied for data treatment. The predi cted results obtained in a test set showed a relative error of 4.3% for suc cinic acid and 5.5% for oxalic acid. (C) 2000 Elsevier Science B.V. All rig hts reserved.