The calibration of ultrasound transducers used to monitor motion of the residual femur within a trans-femoral socket during gait

Citation
Kd. Murray et P. Convery, The calibration of ultrasound transducers used to monitor motion of the residual femur within a trans-femoral socket during gait, PROS ORTHOT, 24(1), 2000, pp. 55-62
Citations number
3
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine
Journal title
PROSTHETICS AND ORTHOTICS INTERNATIONAL
ISSN journal
03093646 → ACNP
Volume
24
Issue
1
Year of publication
2000
Pages
55 - 62
Database
ISI
SICI code
0309-3646(200004)24:1<55:TCOUTU>2.0.ZU;2-X
Abstract
During gait the motion of the residual femur within a trans-femoral socket may be estimated using video recorded data from two ultrasound transducers attached to the socket wall. This paper reviews possible measurement errors and identifies the magnitude of the inaccuracies. Inaccuracies due to equi pment limitations and those due to human error are identified and quantifie d. Ranges of flexion/extension and abduction/adduction of the residual femu r within the socket during gait have been estimated with a cumulative level of inaccuracy of <1 degrees.