Applicability of silicone membrane passive samplers for monitoring of indoor air quality

Citation
B. Zabiegala et al., Applicability of silicone membrane passive samplers for monitoring of indoor air quality, ANAL LETTER, 33(7), 2000, pp. 1361-1372
Citations number
7
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL LETTERS
ISSN journal
00032719 → ACNP
Volume
33
Issue
7
Year of publication
2000
Pages
1361 - 1372
Database
ISI
SICI code
0003-2719(2000)33:7<1361:AOSMPS>2.0.ZU;2-U
Abstract
A series of experiments were conducted to investigate the permeability of m embranes made of silicone foil with respect to selected volatile organic co mpounds such as butanol, benzene, toluene, butyl acetate, ethylbenzene, m-x ylene, styrene, m-dichlorobenzene. On the basis of model experiments the ca libration constants k for 18 passive samplers were determined (with respect to each compound present in the gaseous standard mixture). The obtained calibration constants found are characterised by good precisio n (except for butanol, relative standard deviation are not greater than 9%) . The comparison of the calibration constants for the samplers with silicon e membrane of 50 mu m thick with those obtained for samplers with polyethyl ene membrane of 15.5 mu m shows that silicone membrane samplers will give h igher enrichment factors for the compounds studied.