Direct ICP-OES determination of low concentrations of Cu, Fe, Li, Ti, and V in potassium lithium tantalate niobate crystals

Citation
I. Segal et al., Direct ICP-OES determination of low concentrations of Cu, Fe, Li, Ti, and V in potassium lithium tantalate niobate crystals, ATOM SPECT, 21(2), 2000, pp. 46-49
Citations number
15
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ATOMIC SPECTROSCOPY
ISSN journal
01955373 → ACNP
Volume
21
Issue
2
Year of publication
2000
Pages
46 - 49
Database
ISI
SICI code
0195-5373(200003/04)21:2<46:DIDOLC>2.0.ZU;2-1
Abstract
A fast method of direct trace element determinations in a potassium-tantalu m-niobium matrix has been developed. A new dissolution method was proposed. The PerkinElmer Optima 3300 RL ICP-OES with multicomponent spectral fittin g (MSF) of spectral line corrections was used. Because of the complex matri x, less common spectral lines were chosen. The relative standard deviations (RSDs) for trace element (Cu, Fe, Li, Ti, and V) determinations were 0.2-0 .6%, the limits of detection (LOD) in solutions were 0.005-0.015 mg/L, whic h is equivalent to 5 - 15 ppm in solid samples.