Mm. Kushnir et G. Komaromy-hiller, Optimization and performance of a rapid gas chromatography-mass spectrometry analysis for methylmalonic acid determination in serum and plasma, J CHROMAT B, 741(2), 2000, pp. 231-241
We have developed a rapid and sensitive GC-MS assay for methylmalonic acid
determination in serum and plasma utilizing an anion exchange solid-phase e
xtraction and trimethylsilyl derivatization. Each step of the procedure was
optimized by the experimental design methods to assure the assay reliable
performance. The limit of detection and limit of quantitation were 0.075 an
d 0.1 mu mol/l. The total coefficient of variation for the method was 9.8,
4.4, and 4.6% at the concentration of 0.2, 3.1, and 6.2 mu mol/l methylmalo
nic acid concentration, respectively. The assay was linear up to 9.0 mu mol
/l, and showed good correlation with a reference method. The method has pro
ven to be reliable in routine production, producing clean chromatography, u
nique ion fragments, and consistent ion mass ratio. (C) 2000 Elsevier Scien
ce B.V. All rights reserved.