Determination of alliin in rat plasma by liquid chromatography-electrospray ionization mass spectrometry

Citation
Yg. Shin et H. Matsuura, Determination of alliin in rat plasma by liquid chromatography-electrospray ionization mass spectrometry, J LIQ CHR R, 23(9), 2000, pp. 1331-1338
Citations number
14
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES
ISSN journal
10826076 → ACNP
Volume
23
Issue
9
Year of publication
2000
Pages
1331 - 1338
Database
ISI
SICI code
1082-6076(2000)23:9<1331:DOAIRP>2.0.ZU;2-P
Abstract
A rapid and sensitive high performance liquid chromatography-electrospray/M S method has been developed to determine alliin in rat plasma. The method i nvolved deproteinization of these samples with 2.5-volumes (v/w) of acetoni trile/ethanol (1:1) and then 5 mu L aliquots of the supernatant were inject ed onto a amino-silica (NH2) column coupled with an electrospray/MS system. The mobile phase employed isocratic elution with 5% acetonitrile for 10 mi n; the flow rate was 1.0 mL/min. The column effluent was analyzed by select ed ion monitoring (SIM) for the negative pseudo-molecular ion of alliin [M- H](-) at m/z 176. The limit of detection for alliin in rat plasma by this m ethod was ca. 10 ng/mL and the accuracy for intra- and inter-assay were bet ween 88.3% and 110.1%. The coefficients of variation of the intra- and inte r-assay were generally low (below 9.2% and 11.0%, respectively) for rat pla sma. No problems were encounted due to interference from endogenous substan ces.