Objective: To assess the extent of the variability in insulin delivery over
time, under conditions used in Australian neonatal units, studying the fol
lowing variables: diluent, preconditioning, flushing, sequential preconditi
oning and flushing, insulin concentration, flow rate, catheter type, and ad
dition of albumin.
Methodology: A range of simulated insulin infusions was studied. Infusions
were run over 22 h, with aliquots of infusate collected at baseline and aft
er 15 min, 30 min, 1 h, 2 h, 6 h and 22 h. Insulin concentration was assaye
d using a radioimmunoassay.
Results: An infusion of 50 mU insulin/mL at 1 mL/h gave negligible insulin
delivery until 22 h. However, after preconditioning and flushing the tubing
, consistent insulin delivery was attained by 6 h. An infusion of 200 mU in
sulin/mL at 1 mL/h did not provide consistent delivery until 6 h. At this c
oncentration and rate, consistent insulin delivery was attained within 30 m
in of the start of the infusion either by preconditioning and flushing the
tubing, or by addition of albumin to the infusate.
Conclusion: Clinically significant variation in intravenous insulin deliver
y will occur in the neonatal setting unless counter-measures are taken, suc
h as sequential preconditioning and flushing of the delivery system or the
addition of albumin to the infusate.