Jh. Nurnberg et al., Transthoracic pacing in a very low birth weight infant with congenital complete atrioventricular block, PACE, 23(5), 2000, pp. 907-909
Citations number
13
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
We report our experience of pacemaker treatment in a premature infant of 83
0 g with congenital complete atrioventricular block due to maternal Sjogren
's Syndrome. The infant was delivered by cesarean section at an estimated g
estational age of 26 weeks because of fetal bradycardia, decreasing fetal m
ovements,and hydrops. Immediate postnatal transesophageal ventricular pacin
g was not successful, whereas transthoracic pacing with self-adhesive patch
electrodes adapted to body size resulted in an effective increase of the i
nfant's heart rate until operative application of temporary epimyocardial p
acing wires allowed external stimulation of the heart.