The delivery of 37 shocks by an ICD within 20 minutes, in response to T wav
e oversensing during atrial flutter, resulted in several manifestations of
undesirable device behavior. The generator reverted to backup mode, and dis
abled automatic capacitor reformation, therapy delivery, and automatic gain
control. Postexplant analysis of the device revealed damage to the high vo
ltage output section of the circuitry consistent with excessive electrical
stress. In rare circumstances, multiple infernal discharges can result in s
erious clinical anomalies in ICD behavior, and possibly in an increase in s
usceptibility to circuitry damage.