ICD hardware failure associated with multiple internal shocks

Citation
La. Nair et al., ICD hardware failure associated with multiple internal shocks, PACE, 23(5), 2000, pp. 910-913
Citations number
4
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
23
Issue
5
Year of publication
2000
Pages
910 - 913
Database
ISI
SICI code
0147-8389(200005)23:5<910:IHFAWM>2.0.ZU;2-I
Abstract
The delivery of 37 shocks by an ICD within 20 minutes, in response to T wav e oversensing during atrial flutter, resulted in several manifestations of undesirable device behavior. The generator reverted to backup mode, and dis abled automatic capacitor reformation, therapy delivery, and automatic gain control. Postexplant analysis of the device revealed damage to the high vo ltage output section of the circuitry consistent with excessive electrical stress. In rare circumstances, multiple infernal discharges can result in s erious clinical anomalies in ICD behavior, and possibly in an increase in s usceptibility to circuitry damage.